Reading speed: 266 tests with ISE and 180 tests without ISE
Number of Repositories: 32 Reagent Locations, Used in Cc10-Cc25 and Cc50 Bottles Equipped with Cooling System. Possibility to use any of the stations as R3-R2-R1 reagents
Number of sample placements: 62 positions for the sample and the usability of each position for placement of the calibrator-control-serum routine-neonate and emergency serum. Capable of using 13x75mm tubes and a box equipped with a bar code reader
Reagent System Sampling System: 1000 μl syringe for reagent sampling and 100 μl syringe for sample collection. Removable volume μL 400-110 for R1 and removable volume μl 180 -0 for R2 and R3. Sample removal volume of μl 30-1 with 0.1 μl yield, Reagent needle is equipped with an initial heater. In order to prevent the contamination of the reagents and samples, the prescription needles are equipped with a surface sensor and the presence of the system allows for the detection of the volume in the reference bottles. Reagent and sample needles, in addition to the surface sensor, are also equipped with a collision sensor.
Reaction Tray (Cuvette): The reaction tray has 48 interconnecting cuvettes, complete with automatic and automatic drying system. Peltier element heater for temperature supply of 37 ° C with precision (± 0.1 ° C)
Optical system: 12 watt halogen bulb with 20 watts with 8 optical filters with wavelengths (340 – 405 – 505 – 546 – 578 – 620 – 700 – 660 -), one – to – one
Photometric suffering: Linear photometric suffering -0.1-3.0 with an accuracy of 0.001Abs
Reading methods: Ability to measure with kinetic (bichromatic end point, two point, end point) methods and the ability to plot the automatic reaction dilution process of specific samples (samples whose concentration exceeds the limit of linear readings of the kit). One-point and multi-point calibration (for tests with a curve-shaped reaction), Prozone Check for CUT OFF immutability tests and compute blank tests, reagent blank if necessary.
Quality control: the ability to define 15 different controls and the possibility of placing 3 controls for a monthly quality control charts test and applying the rules of the westgard
Standards: CB, CE-IVD and UL
Computer Specifications and Software: PC with Touch Screen, Windows XP MS Software Included with RS232 Output (Network)
Power consumption and environmental specifications and distilled water consumption: Vac 240-100 volts and maximum power consumption VA380 and 60-60 Hz Frequency can be used at ambient temperature 15-15 ° C and humidity 15-85% Distilled water 0.5 liters per hour
Extras: (According to the order and not available with the device) Bar code reader Isolation of dilute and thick sewage | ISE for measuring Potassium-sodium-Chloride and Co2 | Printer
Dimensions and weight of the device: length 122 cm, width 61 cm, height 75 cm, weight 95 kg